Stimulation Protocols (TMS / tES)

Before you read further

Transcranial stimulation can cause harm, including seizure. Use only under a protocol approved by your ethics board, on a rig whose device-level interlocks are intact, with a trained operator present who can stop stimulation. eeg-mcp is not a medical device, and its hardware backends are generic transports you configure — not validated vendor drivers.

Read Safety in full first.


The shape of every protocol

flowchart LR
    A[list_stim_backends] --> B[attach 'log']
    B --> C[rehearse<br/>full protocol]
    C --> D{timing<br/>correct?}
    D -->|no| C
    D -->|yes| E[configure templates<br/>from device manual]
    E --> F[attach hardware]
    F --> G[arm_stim]
    G --> H[deliver]
    H --> I[disarm_stim]

Steps 1–4 involve no device and no risk. Do all of them before step 5.

Step 1 — See what is available

list_stim_backends()
{
  "hardware_enabled": false,
  "limits": {"max_intensity": 100.0, "max_duration_ms": 10000.0,
             "min_interval_ms": 0.0},
  "backends": [
    {"name": "log", "is_hardware": false, "available": true},
    {"name": "lsl", "is_hardware": false, "available": true},
    {"name": "tms", "is_hardware": true, "available": false,
     "blocked_reason": "EEG_MCP_ALLOW_HARDWARE_STIM is not set"},
    {"name": "tes", "is_hardware": true, "available": false,
     "blocked_reason": "EEG_MCP_ALLOW_HARDWARE_STIM is not set"}
  ]
}

hardware_enabled: false is the default and is correct until an operator deliberately changes it in the server's environment.

Step 2 — Rehearse the whole protocol on log

attach_stim_backend(session_id="s1", backend="log")

The log backend accepts every command and emits nothing. This is not a placeholder — it is how you validate timing, event logging, and your own loop logic with zero risk.

A 10 Hz rTMS train, 20 pulses:

send_stim_train(session_id="s1", label="rtms_10hz", n_pulses=20, interval_ms=100)
{
  "n_delivered": 20,
  "requested_interval_ms": 100.0,
  "realised_interval_ms": {"mean": 100.02, "min": 97.16, "max": 103.20},
  "total_sec": 1.900
}

Read those realised numbers carefully. Mean is accurate and there is no cumulative drift — pulses are scheduled against the train's start, not the previous pulse — but individual gaps vary by a few milliseconds.

Python is not a real-time system

A few ms of jitter is fine for cues and for tES ramps. It is not adequate for sub-millisecond rTMS timing. For that, program the train into the stimulator's own sequencer and use eeg-mcp to arm, gate, and record it.

Step 3 — Get the command strings from your device manual

The hardware backends send what you tell them to. Templates are filled with str.format from the command's fields, so {intensity}, {value}, {duration_ms}, {label} and anything in params are available.

Placeholders are substituted textually — there is no expression evaluation, so a template cannot execute code.

Step 4 — Enable hardware in the server config

This is an operator action, in the MCP client config — not something an agent can do:

{
  "mcpServers": {
    "eeg-realtime": {
      "command": "/path/to/envs/eeg-mcp/bin/python",
      "args": ["-m", "eeg_mcp"],
      "env": {
        "EEG_MCP_ALLOW_HARDWARE_STIM": "1",
        "EEG_MCP_SERIAL_PORT": "COM3",
        "EEG_MCP_MAX_STIM_INTENSITY": "60",
        "EEG_MCP_MIN_STIM_INTERVAL_MS": "95"
      }
    }
  }
}

Set the ceilings to your protocol's approved limits, not to the defaults. The interval floor is what prevents a miscalculated loop turning a 10 Hz train into a 100 Hz one.


rTMS example

Attach

attach_stim_backend(session_id="s1", backend="tms", options={
  "port": "COM3",
  "templates": {
    "set_intensity": "AMP {intensity:.0f}\r\n",
    "fire":  "TRIG\r\n",
    "pulse": "TRIG\r\n"
  },
  "read_reply": true
})
{
  "attached": "tms",
  "status": {"open": true, "armed": false, "is_hardware": true,
             "capabilities": {"intensity_unit": "%MSO"}},
  "next": "Call arm_stim before sending -- this backend drives hardware."
}

Attaching opens the port. It does not permit stimulation.

Arm

arm_stim(session_id="s1", backend="tms", ttl_sec=120)
{"armed": true, "expires_in_sec": 120.0}

Keep the window as short as the block. Arming expires so a stalled agent cannot resume and fire minutes later on a stale decision.

Deliver

Single pulse at 60% MSO:

send_stim_event(session_id="s1", label="tms_single", value=1,
                kind="pulse", intensity=60, unit="%MSO", target="M1_left")
{
  "backend": "tms",
  "is_hardware": true,
  "dispatch_latency_ms": 4.21,
  "detail": {"delivered": true,
             "commands_sent": ["AMP 60\r\n", "TRIG\r\n"]},
  "event": {"label": "tms_single", "origin": "stim"}
}

Intensity is set before firing — the ordering every stimulator protocol assumes, and the one that fails safe if the second step errors.

A 10 Hz train:

send_stim_train(session_id="s1", label="rtms_10hz", n_pulses=20,
                interval_ms=100, intensity=60)

Disarm the moment the block ends

disarm_stim(session_id="s1", backend="tms")

tES example

tES differs in that current is ramped and held for minutes rather than pulsed.

attach_stim_backend(session_id="s1", backend="tes", options={
  "port": "COM4",
  "max_current_ma": 2.0,
  "templates": {
    "ramp":   "RAMP {intensity:.2f} {duration_ms:.0f}\r\n",
    "fire":   "START {intensity:.2f}\r\n",
    "custom": "STOP\r\n"
  }
})

Note max_current_ma: 2.0 — set to the protocol's approved current, below the backend's own 4 mA default ceiling.

arm_stim(session_id="s1", backend="tes", ttl_sec=1500)

send_stim_event(session_id="s1", label="tdcs_ramp_up", kind="ramp",
                intensity=2.0, unit="mA", duration_ms=30000,
                target="F3_anodal")

Long durations need the duration ceiling raised deliberately:

"env": { "EEG_MCP_MAX_STIM_DURATION_MS": "1200000" }

Stop:

send_stim_event(session_id="s1", label="tdcs_stop", kind="custom", intensity=0)
disarm_stim(session_id="s1", backend="tes")

Software targets

Often you want to mark, not stimulate.

LSL — reach other software

attach_stim_backend(session_id="s1", backend="lsl",
                    options={"stream_name": "experiment-markers"})
send_stim_event(session_id="s1", label="cue_left", value=3)

PsychoPy, OpenViBE, LabRecorder and any other LSL client on the machine sees the marker and timestamps it against their own streams. Needs pip install "eeg-mcp[lsl]".

BrainFlow marker channel — sample-aligned

attach_stim_backend(session_id="s1", backend="brainflow_marker")

The highest-fidelity option: BrainFlow interleaves the marker with the samples itself, so no clock join is ever needed. Requires a live BrainFlow stream.

Serial TTL — trigger boxes

attach_stim_backend(session_id="s1", backend="serial_ttl",
                    options={"port": "COM5", "pulse_ms": 10})
send_stim_event(session_id="s1", label="trigger", value=64, kind="pulse")

Writes the code, holds it for pulse_ms, then writes zero — the convention for boxes that latch until cleared.


What gets recorded, always

Every command is logged as an event whether or not the transport succeeded:

get_events(session_id="s1", origin="stim")
{
  "events": [
    {"label": "tms_single", "origin": "stim", "value": 1.0,
     "sample_index": 128400,
     "payload": {"backend": "tms", "delivered": true, "marked_in_source": true,
                 "commands_sent": ["AMP 60\r\n", "TRIG\r\n"]}},
    {"label": "tms_single", "origin": "stim",
     "payload": {"backend": "tms", "delivered": false,
                 "error": "Only 42.1 ms since the last stimulation; the minimum
                           safe interval is 95.0 ms."}}
  ]
}

A refused stimulation is as much a part of the record as a delivered one. Combined with the EEG on the same clock, the session reconstructs completely.

Troubleshooting

Error Meaning Fix
Backend 'tms' ... is disabled Config gate Operator sets EEG_MCP_ALLOW_HARDWARE_STIM=1
not armed (or arming expired) Arm gate arm_stim again
intensity=85 exceeds the configured ceiling Limit gate Check the protocol; raise the ceiling only if genuinely required
Only 42.1 ms since the last stimulation Interval floor Your loop is firing too fast
No 'fire' template configured Missing template Add it from the device manual
Template 'fire' references 'freq' Unknown placeholder Pass it via params={"freq": 10}